  • Treatment Power: 75 W PIP
  • Sample Processing Capacity: Single Tube
  • NGS Sample Volumes: 15 to 130 µl

The M220 Focused-ultrasonicator offers the “Scientist’s Standard” in a compact, easy-to-use system making it the ideal DNA shearing solution for MiSeq and PGM users. AFA® technology in the M220 eliminates operator-induced variations, improves performance, increases efficiency, and provides standardized results.

More Information
Technical Specification Description
Primary Applications Cell lysis: mammalian, bacterial, yeast, cfDNA Extraction from Plasma, DNA Extraction from Dried Blood Spots, DNA/RNA/Chromatin Shearing, DNA/RNA/Total NA/Protein Extraction from FFPE, Tissue lysis and extraction
Treatment Power 75 W PIP
Part Number 500295
Power Requirements 100-240 VAC 500 VA, 50-60Hz
Operating Environment Ambient temperature: 19 °C to 25 °C (66 °F to 77 °F) Relative humidity: 30% to 70%
Operating Temperature Internal temperature control: Protocol temperature range from 7 °C to 20 °C
Safety Complies with Low Voltage Directive 2006/95/EC. Certified to IEC/EN/UL 61010-1:2004 “Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use, Part1: General Requirements"
Regulatory Labeling CE, ETL Mark (for Product Safety), WEEE
SonoLab Version SonoLab 7
Operating System Windows 10
WCS NA
Chiller Solid state chiller built-in
Sample Processing Capacity Single Tube
NGS Sample Volumes 15 to 130 µl
Recommended Batch Size NA
Covaris Consumables microTUBE, microTUBE-50, milliTUBE 1 ml, miniTUBE
Upgrade Potential No
Associated Accessories M220 Accessories

Key Features

  • Controlled energy delivery
  • Isothermal processing
  • Non-contact, closed vessel
  • Quiet operation
  • Robust and reliable design
  • Tried and tested in major genome centers worldwide

Benefits

  • User-selectable fragment lengths
  • No heat damage, high sample recovery
  • No cross contamination, no aerosols, no clean up
  • No discomfort to operators
  • Reproducible results, year after year
  • The mechanical shearing method of choice

DNA Shearing Protocols for Ion Torrent Platforms with the M220 Focused-ultrasonicator

Download

Quick Guide: DNA Shearing with M220 Focused-ultrasonicator

Download

M220 Focused-ultrasonicator Setup & Instruction Manual

Product Inserts and User Manuals

Download

Quick Start Guide: M220 Focused-ultrasonicator Instrument

Product Inserts and User Manuals

Download

Formulation and Drug Delivery Breakthrough processing for poorly soluble API’s

Brochures

Download

Covaris-certified Consumable Guide

Brochures

Download

DNA Shearing for Next Generation Sequencing (NGS) with the M220 Focused-ultrasonicator

DNA Shearing for NGS with the ME220 Focused-ultrasonicator.

Download

Magnetic Bead-based Clean-up Using AFA-energetics®

When precisely tuned, this stream of Covaris Adaptive Focused Acoustic energy enables DNA shearing precisely and reliably to desired fragment sizes, in addition to gentle...

Download

Protein Extraction from Yeast: Comparison of the Covaris Adaptive Focused Acoustics®(AFA) Process to Conventional Bead Beating and Probe Sonication

The efficiency of several mechanical-based lysis and extraction techniques, such as Adaptive Focused Acoustics (AFA), probe sonication, and bead beating from yeast isolates was compared.

Download