Quick Guide: DNA Shearing with E220 Focused-ultrasonicator
Download
The E220evolution enables multi-sample, batch preparation, capable of processing a wide range of sample types and volumes and may be programed to process from 1 to 8 samples in a single batch. When workload increases, the E220evolution can be upgraded to an E220 capable of processing 1-96 samples/run or to an LE220 for high-throughput processing without purchasing a new instrument.
| Technical Specification | Description |
|---|---|
| Primary Applications | Cell lysis: mammalian, bacterial, yeast, cfDNA Extraction from Plasma, DNA Extraction from Dried Blood Spots, DNA/RNA/Chromatin Shearing, DNA/RNA/Total NA/Protein Extraction from FFPE, Tissue lysis and extraction |
| Treatment Power | 100 W Avg. Incident Power |
| Part Number | 500429 |
| Power Requirements | E100-120, 200-240 V~ 50-60Hz, 500 VAC |
| Operating Environment | Ambient temperature: 19 °C to 25 °C (66 °F to 77 °F) Relative humidity: 30% to 70% |
| Operating Temperature | 0.5 °C to 55.0 °C |
| Safety | Complies with Low Voltage Directive 2006/95/EC. Certified to IEC/EN/UL 61010-1:2004 “Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use, Part1: General Requirements |
| Regulatory Labeling | CE and ETL Mark |
| SonoLab Version | SonoLab 7 |
| Operating System | Windows 10 |
| WCS | Yes |
| Chiller | Chiller/heater re-circulating system-Not included, sold separately |
| Sample Processing Capacity | Single Tube |
| NGS Sample Volumes | 15 to 130 µl |
| Recommended Batch Size | 1 to 8 Samples |
| Covaris Consumables | microTUBE 8-strip (rack required) |
| Upgrade Potential | Yes |
| Associated Accessories | E220evolution Accessories |
Key Features
Benefits
Quick Guide: DNA Shearing with E220 Focused-ultrasonicator
Download